首页 /研究 /In Situ Flexible Needle Adjustment Towards MRI-Guided Spinal Injections Based on Finite Element Simulation
MANIPULATION

In Situ Flexible Needle Adjustment Towards MRI-Guided Spinal Injections Based on Finite Element Simulation

Yanzhou Wang, Yangsheng Xu, Ka‐Wai Kwok, Iulian Iordachita

发表年份
2023
引用次数
3

摘要

This paper investigates the possibility of roboti-cally performing in situ needle manipulations to correct the needle tip position in the setting of robot-assisted, MRI -guided spinal injections, where real time MRI images cannot be effectively used to guide the needle. Open-loop control of the needle tip is derived from finite element simulation, and the proposed method is tested with ex vivo animal muscle tissues and validated by cone beam computed tomography. Preliminary results have shown promise of performing needle tip correction in situ to improve needle insertion accuracy when real-time feedback is not readily available.

关键词

Finite element methodIn situBiomedical engineeringComputer scienceRobotMagnetic resonance imagingMaterials scienceArtificial intelligenceEngineeringRadiology

相关论文