首页 /研究 /Multi-step prediction of physiological tremor for robotics applications
SURGICAL

Multi-step prediction of physiological tremor for robotics applications

Kalyana C. Veluvolu, Sivanagaraja Tatinati, Seung‐Mo Hong, Wei Tech Ang

发表年份
2013
引用次数
8

摘要

The performance of surgical robotic devices in real-time mainly depends on phase-delay in sensors and filtering process. A phase delay of 16-20 ms is unavoidable in these robotics procedures due to the presence of hardware low pass filter in sensors and pre-filtering required in later stages of cancellation. To overcome this phase delay, we employ multi-step prediction with band limited multiple Fourier linear combiner (BMFLC) and Autoregressive (AR) methods. Results show that the overall accuracy is improved by 60% for tremor estimation compared to single-step prediction methods in the presence of phase delay. Experimental results with the proposed methods for 1-DOF tremor estimation highlight the improvement.

关键词

Computer scienceRoboticsAutoregressive modelArtificial intelligenceProcess (computing)Filter (signal processing)Group delay and phase delayPhase (matter)RobotComputer vision

相关论文

查看 SURGICAL 分类全部论文