首页 /研究 /Shoulder-mounted Robot for MRI-Guided Arthrography: Clinically Optimized System
OTHER

Shoulder-mounted Robot for MRI-Guided Arthrography: Clinically Optimized System

Gyeong Hu Kim, Niravkumar Patel, Jiawen Yan, Di Wu, Gang Li, Kevin Cleary, Iulian Iordachita

发表年份
2019
引用次数
10

摘要

This paper introduces our compact and lightweight patient-mounted MRI-compatible 4 degree-of-freedom (DOF) robot with an improved transmission system for MRI-guided arthrography procedures. This robot could make the traditional two-stage arthrography procedure (fluoroscopy-guided needle insertion followed by a diagnostic MRI scan) simpler by converting it to a one-stage procedure but more accurate with an optimized system. The new transmission system is proposed, using different mechanical components, to result in higher accuracy of needle insertion. The results of a recent accuracy study are reported. Experimental results show that the new system has an error of 1.7 mm in positioning the needle tip at a depth of 50 mm, which indicates high accuracy.

关键词

FluoroscopyComputer scienceRobotTransmission (telecommunications)Computer visionArtificial intelligenceBiomedical engineeringRadiologyMedicine

相关论文

查看 OTHER 分类全部论文