Shoulder-mounted Robot for MRI-Guided Arthrography: Clinically Optimized System
Gyeong Hu Kim, Niravkumar Patel, Jiawen Yan, Di Wu, Gang Li, Kevin Cleary, Iulian Iordachita
- Year
- 2019
- Citations
- 10
Abstract
This paper introduces our compact and lightweight patient-mounted MRI-compatible 4 degree-of-freedom (DOF) robot with an improved transmission system for MRI-guided arthrography procedures. This robot could make the traditional two-stage arthrography procedure (fluoroscopy-guided needle insertion followed by a diagnostic MRI scan) simpler by converting it to a one-stage procedure but more accurate with an optimized system. The new transmission system is proposed, using different mechanical components, to result in higher accuracy of needle insertion. The results of a recent accuracy study are reported. Experimental results show that the new system has an error of 1.7 mm in positioning the needle tip at a depth of 50 mm, which indicates high accuracy.
Keywords
Related papers
Statistical Learning Theory
Yuhai Wu, Vladimir Vapnik
1999
Artificial intelligence: a modern approach
1995
Fractional Differential Equations
Igor Podlubný
2025
Applied Nonlinear Control
Jean-Jacques Slotine, Weiping Li
1991