首页 /研究 /Automating multi-throw multilateral surgical suturing with a mechanical needle guide and sequential convex optimization
SURGICAL

Automating multi-throw multilateral surgical suturing with a mechanical needle guide and sequential convex optimization

Siddarth Sen, Animesh Garg, David V. Gealy, Stephen McKinley, Yiming Jen, Ken Goldberg

发表年份
2016
引用次数
168

摘要

For supervised automation of multi-throw suturing in Robot-Assisted Minimally Invasive Surgery, we present a novel mechanical needle guide and a framework for optimizing needle size, trajectory, and control parameters using sequential convex programming. The Suture Needle Angular Positioner (SNAP) results in a 3x error reduction in the needle pose estimate in comparison with the standard actuator. We evaluate the algorithm and SNAP on a da Vinci Research Kit using tissue phantoms and compare completion time with that of humans from the JIGSAWS dataset [5]. Initial results suggest that the dVRK can perform suturing at 30% of human speed while completing 86% suture throws attempted. Videos and data are available at: berkeleyautomation.github.io/amts.

关键词

Computer scienceAutomationRobotFibrous jointTrajectoryConvex optimizationReduction (mathematics)Haptic technologyRegular polygonArtificial intelligence

相关论文