Home /Research /A statically balanced and bi-stable compliant end effector combined with a laparoscopic 2DoF robotic arm
SURGICAL

A statically balanced and bi-stable compliant end effector combined with a laparoscopic 2DoF robotic arm

J. Lassooij, N. Tolou, Giuseppe Tortora, Sebastiano Caccavaro, Arianna Menciassi, Just L. Herder

Year
2012
Citations
27
Access
Open access

Abstract

Abstract. This article presents the design of a newly developed 2DoF robotic arm with a novel statically balanced and bi-stable compliant grasper as the end effector for laparoscopic surgery application. The arm is based on internal motors actuating 2 rotational DoFs: pitch and roll. The positive stiffness of the monolithic grasper has been compensated using pre-curved straight guided beams that are preloaded collinear with the direction of actuation of the grasper. The result is a fully compliant statically balanced laparoscopic grasper. The grasper has been successfully adapted to a robotic arm. The maximum force and stiffness compensations were measured to be 94% and 97% (i.e. near zero stiffness) respectively. Furthermore, the feasibility of adjusting for bi-stable behavior has been shown. This research can be a preliminary step towards the design of a statically balanced fully compliant robotic arm for laparoscopic surgery and similar areas.

Keywords

StiffnessCompliant mechanismRobotic armRobot end effectorEngineeringControl theory (sociology)Computer scienceStructural engineeringSimulationRobot

Related papers

Browse all SURGICAL papers