首页 /研究 /Development and Preliminary Evaluation of a Robotic Device for MRI-Guided Needle Breast Biopsy
OTHER

Development and Preliminary Evaluation of a Robotic Device for MRI-Guided Needle Breast Biopsy

Cleanthis Ioannides, Anastasia Antoniou, Vasiliki Zinonos, Christakis Damianou

发表年份
2024
引用次数
2

摘要

This study concerns the development and evaluation of a simple and ergonomic robotic system for Magnetic Resonance Imaging (MRI)-guided needle breast biopsy with lateral needle approach. The device comprises two piezoelectrically actuated linear motion stages intended to align a needle supporter with the target for manual needle insertion. The device demonstrated submillimeter accuracy and safe operation within a 3 T clinical MRI scanner. In phantom studies, tumor simulators of varying sizes were successfully targeted in both laboratory and MRI settings.

关键词

Breast biopsyBiopsyMedicineBreast MRIRadiologyMedical physicsBiomedical engineeringBreast cancerMammographyInternal medicine