Home /Research /Development and Preliminary Evaluation of a Robotic Device for MRI-Guided Needle Breast Biopsy
OTHER

Development and Preliminary Evaluation of a Robotic Device for MRI-Guided Needle Breast Biopsy

Cleanthis Ioannides, Anastasia Antoniou, Vasiliki Zinonos, Christakis Damianou

Year
2024
Citations
2

Abstract

This study concerns the development and evaluation of a simple and ergonomic robotic system for Magnetic Resonance Imaging (MRI)-guided needle breast biopsy with lateral needle approach. The device comprises two piezoelectrically actuated linear motion stages intended to align a needle supporter with the target for manual needle insertion. The device demonstrated submillimeter accuracy and safe operation within a 3 T clinical MRI scanner. In phantom studies, tumor simulators of varying sizes were successfully targeted in both laboratory and MRI settings.

Keywords

Breast biopsyBiopsyMedicineBreast MRIRadiologyMedical physicsBiomedical engineeringBreast cancerMammographyInternal medicine

Related papers

Browse all OTHER papers