首页 /研究 /Robot-assisted navigation for percutaneous localization of peripheral pulmonary nodule: an in vivo swine study
SURGICAL

Robot-assisted navigation for percutaneous localization of peripheral pulmonary nodule: an in vivo swine study

Xingguang Duan, Rui He, Yu Jiang, Fei Cui, Hao Wen, Xiangqian Chen, Zhexue Hao, Yuan Zeng, Hui Liu, Jipeng Shi, Houiam Cheong, Mengxing Dong, U Kaicheng, Shunjun Jiang, Wei Wang, Hengrui Liang, Jun Liu, Jianxing He

发表年份
2023
引用次数
8
访问权限
开放获取

摘要

Background: Robot-assisted surgery (RAS) systems have been developed but rarely applied to lung nodule localization. This study aimed to assess the feasibility and safety of using a robot-assisted navigation system in percutaneous lung nodule localization. Methods: A computed tomography (CT)-guided robot-assisted navigation system was used to localize the simulated peripheral nodule in the swine lung through fluorescent agent injection. After the localization, fluorescent thoracoscopic wedge resection was performed. The deviation between the target point and the needle tip was measured using a professional 3-dimensional (3D) distance measurement software. The primary outcome was the localization accuracy (deviation) of the localization. The secondary outcomes were the localization-related complication rate, the localization duration, and the success rate. Results: A total of 4 pigs were enrolled, and 20 peripheral lung nodules were created and localized successfully. All nodules underwent subsequent wedge resection for verification. The mean deviation by measuring the 3D distance was 3.81 mm [standard deviation (SD): 1.29 mm, 95% confidence interval (CI): 2.936-4.536 mm]. The technical success rate for localization was 100%, and the mean localization time was 14.69 minutes (SD: 4.67 minutes). The complication rate was 5% (1/20), with 1 pneumothorax after localization, and no mortality occurred. Conclusions: This pilot animal study demonstrated the promising potential of the robot-assisted navigation technique in peripheral lung nodule localization, with high accuracy and feasibility. Further clinical trials are needed to validate its safety compared to traditional manual localization.

关键词

PercutaneousNodule (geology)MedicinePneumothoraxLungWedge resectionConfidence intervalPeripheralRadiologyResection

相关论文

查看 SURGICAL 分类全部论文