首页 /研究 /A novel master–slave teleoperation robot system for diaphyseal fracture reduction: a preliminary study
HRI

A novel master–slave teleoperation robot system for diaphyseal fracture reduction: a preliminary study

Changsheng Li, Tianmiao Wang, Lei Hu, Peifu Tang, Lifeng Wang, Lihai Zhang, Na Guo, Yiming Tan

发表年份
2016
引用次数
31
访问权限
开放获取

摘要

We developed a novel master–slave teleoperation robot system for diaphyseal fracture reduction. This paper details the mechanical design and the control system as a preliminary study. The fracture reduction accuracy and the performance of the control system are experimentally tested. The mean error in the axial and lateral displacements was below 3 mm; the mean side angle and inward turn errors were below 4°. As a preliminary study, the feasibility of the novel master–slave teleoperation robot system for diaphyseal fracture reduction is verified. The robot system fulfills the reduction requirements in accuracy and operating performance.

关键词

TeleoperationReduction (mathematics)RobotSimulationMaster/slaveFracture (geology)Computer scienceControl systemControl theory (sociology)Engineering

相关论文

查看 HRI 分类全部论文