Home /Research /Development of a novel robotic platform with controllable stiffness manipulation arms for laparoendoscopic single‐site surgery (LESS)
SURGICAL

Development of a novel robotic platform with controllable stiffness manipulation arms for laparoendoscopic single‐site surgery (LESS)

Jianchen Wang, Shuxin Wang, Jinhua Li, Xiangyun Ren, Randall Briggs

Year
2017
Citations
52

Abstract

BACKGROUND: For current LESS robotic systems, the trade-off between dexterity and payload capability is always present. This paper presents a novel LESS robotic platform equipped with controllable stiffness manipulation arms. METHODS: Each manipulation arm with an articulated section and a controllable stiffness continuum section (CSCS) can be switched between a 7-DoF compliant status and 5-DoF rigid status according to the operation requirement. Screw theory and product exponential formula are used to quantify the kinematic performance. RESULTS: The stiffness of the manipulation arm promotes 3.03 to 4.12 times from compliant to rigid CSCS with maximum payload of 10 N in rigid status. The shortest rigid/compliant switching time is 5 s. The precision of a tracking test and an ex vivo procedure verified the accuracy and effectiveness of the controllable stiffness manipulation arms. CONCLUSIONS: This robot could potentially improve the surgical performance and further expand robotic LESS procedures.

Keywords

Computer scienceRobotic surgeryStiffnessSurgeryMedicineArtificial intelligenceMaterials science

Related papers