A 3D-printed phantom for stereotactic body radiation therapy simulation
Ying-Hao Yu, Tsung-Yu Yen, Shih‐Kai Hung, S.Y. James Chen, Kuei-Ying Wang
- 发表年份
- 2024
- 引用次数
- 2
- 访问权限
- 开放获取
摘要
In modern radiation therapy for lung cancer, examining the uncertainty between tumor motion and beam delivery is vitally important. To lower the radiation dose delivery to the patient's normal tissue, narrowing the irradiation field margin to hit the tumor accurately is critical. Thus we proposed a phantom that simulates the thorax and lung tumor's motions by employing a 3D printing technique. The lung tumor is controlled by a linear miniature Delta robot arm, with a maximum displacement of 20 mm in each direction. When we simulated the thoracic breathing movements at 12 mm in A-P (Anterior-Posterior), the control errors were within 10%. The average tracking errors of the prosthetic tumor were within 1.1 mm. Therefore, the 3D-printed phantom with a robot arm can provide a reliable simulation for training and dosimetry measurement before lung radiotherapy, especially SBRT.