首页 /研究 /Evaluating the resistant force of an endoscopic capsule self-propelling in the small intestine
OTHER

Evaluating the resistant force of an endoscopic capsule self-propelling in the small intestine

Yao Yan, Bingyong Guo, Jiyuan Tian, Jiajia Zhang, Baoquan Zhang, Ellis Ley, Yang Liu, Shyam Prasad

发表年份
2022
引用次数
7
访问权限
开放获取

摘要

Abstract In this paper, we study a mathematical model for evaluating the resistant force of endoscopic devices self-propelling in the small intestine with a consideration of its anatomy. Circular fold is the main source of intestinal resistance that needs to be overcome during the endoscopic procedure. Our model is able to calculate the resistances of such folds in different dimensions. Finite element analysis and experimental testing are presented to validate the proposed model by using a pulling-type endoscopic capsule. Our investigation shows that the resistance reaches its maximum immediately after the capsule is pulled against the fold, and drops off gradually during the crossing motion. The proposed method is further demonstrated by using a magnetic pulling-type capsule prototype in an intestine simulator. The findings of this study provide a better insight into the biomechanics of the small intestine and advance the understanding of capsule–intestine interaction for robotic and endoscopic engineers.

关键词

CapsuleSmall intestineBiomechanicsLarge intestineFinite element methodBiomedical engineeringComputer scienceAnatomySimulationMechanics