Design of a Partially Compliant, Three-Phalanx Underactuated Prosthetic Finger
Marco W. M. Groenewegen, Milton E. Aguirre, Just L. Herder
- 发表年份
- 2015
- 引用次数
- 13
摘要
Advanced robotic hand prostheses are praised for their impressive robust and fine grasping capabilities generated from intricate systems. Nevertheless, a high demand remains for grasping mechanisms that are mechanically simple, lightweight, and cheap to produce, easy to assemble and low in maintenance costs. This paper presents the design of a partially compliant underactuated finger to demonstrate the feasibility of achieving these rigorous requirements. The conceptual topology of the three phalanx finger is selected based on competitive analysis. Employing Pseudo-Rigid Body Model and Finite Element Analysis, a genetic optimization problem is formulated to minimize bending stresses within compliant flexures. The result is a fully functional demonstrator capable of flexing 180° in finger rotation. The prototype is fabricated from flexible high strength nylon and requires no assembly steps beyond 3D printing. Experimental testing verifies the design method with an acceptable error of < 5%.
关键词
相关论文
Statistical Learning Theory
Yuhai Wu, Vladimir Vapnik
1999
Artificial intelligence: a modern approach
1995
Applied Nonlinear Control
Jean-Jacques Slotine, Weiping Li
1991
A new optimizer using particle swarm theory
R.C. Eberhart, James Kennedy
2002