Evaluating the Feasibility of Magnetic Tools for the Minimum Dynamic Requirements of Microneurosurgery
Cameron Forbrigger, Erik Fredin, Eric Diller
- 发表年份
- 2023
- 引用次数
- 13
摘要
Neurosurgery could benefit from robot-assisted minimally invasive approaches, but existing robot tools are insufficiently small and compact. Magnetic actuation is an attractive approach to medical robotics because it allows small, modular serial mechanisms to be remotely actuated. Despite these advantages, magnetic actuation is relatively weak compared to alternative actuation methods. In this paper, we introduce a novel analytical model for magnetic serial robots, use this model to design two prototypes, and then demonstrate that a 4-mm-diameter prototype without any internal mechanical transmission can produce forces up to 0.181 N: high enough to perform delicate microsurgical tasks. We also demonstrate that the robot can achieve a closed-loop step response rise time of 0.71 seconds with an overshoot of 7.8%: sufficiently fast for surgical motions while maintaining a tip precision of less than 2 mm during a worst-case dynamic motion. These experiments provide strong evidence for the feasibility of directly-driven magnetic tools for neurosurgical applications, and they motivate future investigations in this area.
关键词
相关论文
Statistical Learning Theory
Yuhai Wu, Vladimir Vapnik
1999
Artificial intelligence: a modern approach
1995
Applied Nonlinear Control
Jean-Jacques Slotine, Weiping Li
1991
A new optimizer using particle swarm theory
R.C. Eberhart, James Kennedy
2002