Magnetic resonance image–guided focused ultrasound robotic system for transrectal prostate cancer therapy

Marinos Giannakou, Theocharis Drakos, Georgios Menikou, Nikolas Evripidou, Antria Filippou, Kyriakos Spanoudes, Leonidas Ioannou, Christakis Damianou

发表年份
2021
引用次数
24
访问权限
开放获取

摘要

BACKGROUND: A magnetic resonance image (MRI) guided robotic device for focussed ultrasound therapy of prostate cancer (PC) was developed. The device offers movement in 5 degrees of freedom (DOF) and uses a single-element transducer that operates at 3.2 MHz, has a diameter of 25 mm and focuses at 45 mm. METHODS: The MRI compatibility of the system was evaluated in a 1.5 T scanner. The ability of the transducer to create lesions was evaluated in laboratory and MRI settings, on ex vivo pork tissue and in vivo rabbit thigh tissue. RESULTS: Cavitational and thermal lesions were created on the excised pork tissue. In vivo experiments proved the efficacy of the system in ablating muscle tissue without damaging intervening areas. CONCLUSIONS: The MRI compatible robotic system can be placed on the table of any commercial MRI scanner up to 7 T. The device has the ability of future use for transrectal focal therapy of PC with the patient in supine position.

关键词

Magnetic resonance imagingProstateUltrasoundSupine positionMedicineProstate cancerBiomedical engineeringScannerInterventional magnetic resonance imagingTransducer