Magnetic resonance image–guided focused ultrasound robotic system for transrectal prostate cancer therapy
Marinos Giannakou, Theocharis Drakos, Georgios Menikou, Nikolas Evripidou, Antria Filippou, Kyriakos Spanoudes, Leonidas Ioannou, Christakis Damianou
- Year
- 2021
- Citations
- 24
- Access
- Open access
Abstract
BACKGROUND: A magnetic resonance image (MRI) guided robotic device for focussed ultrasound therapy of prostate cancer (PC) was developed. The device offers movement in 5 degrees of freedom (DOF) and uses a single-element transducer that operates at 3.2 MHz, has a diameter of 25 mm and focuses at 45 mm. METHODS: The MRI compatibility of the system was evaluated in a 1.5 T scanner. The ability of the transducer to create lesions was evaluated in laboratory and MRI settings, on ex vivo pork tissue and in vivo rabbit thigh tissue. RESULTS: Cavitational and thermal lesions were created on the excised pork tissue. In vivo experiments proved the efficacy of the system in ablating muscle tissue without damaging intervening areas. CONCLUSIONS: The MRI compatible robotic system can be placed on the table of any commercial MRI scanner up to 7 T. The device has the ability of future use for transrectal focal therapy of PC with the patient in supine position.