Automated Needle Targeting with X-ray (ANT-X) - Robot-assisted device for percutaneous nephrolithotomy (PCNL) with its first successful use in human.
Mon Mon Oo, Himesh Ramesh Gandhi, Kian Tai Chong, Ka Wei Ng, Aung Thaw Hein, Yung K. Tan
- 发表年份
- 2018
- 引用次数
- 39
摘要
INTRODUCTION: To make percutaneous access easier in PCNL, we developed Automated Needle Targeting with X-ray (ANT-X). METHOD: ANT-X uses an image registration software with a closed loop feedback system to autoalign the puncture needle to the desired calyx using the bullseye technique. We tried percutaneous punctures on a live pig model and compared the results with free-hand technique. We then performed our first PCNL in a human subject with the aid of ANT-X. Our patient was a 48 year-old gentleman with a 1.4cm left lower pole stone. RESULTS: Initial results for live animal trial showed radiation exposure for robot-assisted arm during puncture was reduced by 26% compared to the free-hand technique (8.2mGy vs 11.2mGy). In the human trial, obtaining percutaneous access was successful at first attempt. CONCLUSION: ANT-X system can help surgeons feel confident and potentially reduce complications, hence enabling more surgeons to adopt this procedure.