Home /Research /Design of a Four-axis Force/Moment Sensor for Measuring the Applied Force to Wrist
OTHER

Design of a Four-axis Force/Moment Sensor for Measuring the Applied Force to Wrist

Tae-Kyung Hong, Gab-Soon Kim

Year
2013
Citations
3
Access
Open access

Abstract

Patients have the paralysis of their wrists, and can't use of their wrists freely. But their wrists can be recovered by wrist-bending rehabilitation exercise. Professional rehabilitation therapeutists exercise the wrists of patients in hospital. But the wrists of patients have not exercised enough for the rehabilitation, because the therapeutists are much less than patients in number. Therefore, the wrist rehabilitation robot should be developed, and it have to measure the applied force to the patients' wrists for their safety. In this paper, the four-axis force/moment sensor was designed for the wrist rehabilitation robot. As a test results, the interference error of the four-axis force/moment sensor was less than 0.91%. It is thought that the sensor can be used to measure the applied force to the patients' wrists.

Keywords

RehabilitationWristPhysical medicine and rehabilitationMoment (physics)MedicinePhysical therapyBraceRobotSimulationComputer science

Related papers

Browse all OTHER papers