Design of MR-compatible robotic devices: magnetic and geometric compatibility aspects
Christoforos Keroglou, Nikolaos V. Tsekos, Ioannis Seimenis, Eleni Eracleous, Christodoulos G. Christodoulou, Eftychios G. Christoforou
- Year
- 2009
- Citations
- 4
Abstract
Specially designed robotic manipulators have been proposed for the performance of minimally invasive interventions under real-time magnetic resonance imaging (MRI) guidance. The design of MR-compatible robotic systems is a challenging task given the limitations imposed by the magnetic nature of the scanning environment but also the geometry of a high-field cylindrical scanner. These issues are discussed with special emphasis on geometric MR-compatibility. Acquired MR images are used for the analysis of the available space inside the scanner in order to provide the necessary input for the design of interventional devices.