Accuracy of the Imageless Mode of the ROSA Robotic System for Targeted Resection Thickness in Total Knee Arthroplasty: A Prospective, Single Surgeon Case‐Series Study
Zakareya Gamie, Eustathios Kenanidis, Georgios Douvlis, Nikolaos Mylonakis, Alexander Maslaris, Eleftherios Tsiridis
- Year
- 2024
- Citations
- 12
- Access
- Open access
Abstract
BACKGROUND: We investigated the accuracy of targeted resection thickness in patients undergoing primary Total Knee Arthroplasty (TKA) using the ROSA robotic system. METHODS: Calliper measurements of the distal femur (DF), proximal tibia (PT), and posterior condyles (PC) were taken in 44 patients from June 2023 to January 2024. RESULTS: Planned and actual resection depth difference was 0.67 mm ± 0.6 mm (mean ± SD) (p = 0.217) and 0.94 mm ± 1.15 mm (p = 0.4) for medial and lateral DF, 0.93 mm ± 0.81 mm (p = 0.001) and 0.89 mm ± 0.8 mm (p = 0.008) for medial and lateral PT, and 1.1 mm ± 0.97 mm (p = 0.001) and 1.04 mm ± 0.79 mm (p = 0.001) for medial and lateral PC, respectively. CONCLUSION: The ROSA robotic system can achieve a high degree of accuracy for planned resection thickness. Results are valid only for the imageless ROSA TKA in patients with primary knee osteoarthritis.
Keywords
Related papers
Statistical Learning Theory
Yuhai Wu, Vladimir Vapnik
1999
Fractional Differential Equations
Igor Podlubný
2025
Applied Nonlinear Control
Jean-Jacques Slotine, Weiping Li
1991
Genetic Programming: On the Programming of Computers by Means of Natural Selection
John R. Koza
1992