Accuracy of the Imageless Mode of the ROSA Robotic System for Targeted Resection Thickness in Total Knee Arthroplasty: A Prospective, Single Surgeon Case‐Series Study
Zakareya Gamie, Eustathios Kenanidis, Georgios Douvlis, Nikolaos Mylonakis, Alexander Maslaris, Eleftherios Tsiridis
- 发表年份
- 2024
- 引用次数
- 12
- 访问权限
- 开放获取
摘要
BACKGROUND: We investigated the accuracy of targeted resection thickness in patients undergoing primary Total Knee Arthroplasty (TKA) using the ROSA robotic system. METHODS: Calliper measurements of the distal femur (DF), proximal tibia (PT), and posterior condyles (PC) were taken in 44 patients from June 2023 to January 2024. RESULTS: Planned and actual resection depth difference was 0.67 mm ± 0.6 mm (mean ± SD) (p = 0.217) and 0.94 mm ± 1.15 mm (p = 0.4) for medial and lateral DF, 0.93 mm ± 0.81 mm (p = 0.001) and 0.89 mm ± 0.8 mm (p = 0.008) for medial and lateral PT, and 1.1 mm ± 0.97 mm (p = 0.001) and 1.04 mm ± 0.79 mm (p = 0.001) for medial and lateral PC, respectively. CONCLUSION: The ROSA robotic system can achieve a high degree of accuracy for planned resection thickness. Results are valid only for the imageless ROSA TKA in patients with primary knee osteoarthritis.