Safety of a 1064‐nm robotic laser system for noninvasive lipolysis of the flanks

Thomas Fiala, Philip T. Lavin

发表年份
2022
引用次数
2

摘要

OBJECTIVES: The primary objective of this pilot study was to confirm the safety of a 1064-nm laser device with a novel robotic arm for noninvasive subcutaneous fat reduction in the flank area. Secondary objectives included: assessing the extent of subject discomfort during treatment, overall subject satisfaction with the results of the procedure, and a determination of subcutaneous fat reduction in the treated area, in preparation for larger upcoming trials. MATERIALS AND METHODS: area on both flanks of enrolled subjects (N = 11; 22 flanks) was treated for 20 minutes with a Food and Drug Administration-cleared robotic noncontact 1064-nm laser system (EON®; Dominion Aesthetic Technologies, Inc.). Patients were followed for 12 weeks, and examined routinely at 2 weeks, 12 weeks, and additionally as needed, posttreatment. Ad hoc surveys were administered to assess patient satisfaction. A 2-week posttreatment ultrasound scan was used to check for changes in the treated area. Ultrasound measurements were also used to determine the subcutaneous adipose tissue thickness at a center of each treatment zone before treatment and at 12 weeks posttreatment for efficacy determination, with mean thicknesses calculated per subject. RESULTS: The treatment had a low incidence of adverse effects, with only one subject developing a palpable thickening in the subcutaneous tissue following treatment. This was noted at the 2-week time period and had resolved by the 12-week posttreatment exam. No other predefined adverse effects were noted. On a scale of 0-10, the mean pain score during the procedure was 1.95, decreasing to 0.9 at 30 minutes postprocedure. Subject satisfaction was "Excellent" for all subjects (100%). At Week 12 after one treatment, the mean reduction in subcutaneous adipose thickness on the treated flanks was 6.1 mm per patient (-15%; p < 0.01). CONCLUSION: Similar to a prior abdominal study with the same robotic laser device, this pilot study confirms the safety of this 1064-nm noncontact laser device for treating subcutaneous fat on the flanks. The procedure is well tolerated with a high degree of subject satisfaction. The amount of subcutaneous fat reduction in the flank area appears similar to that seen in the abdomen, but larger studies are required for confirmation. CLINICALTRIALS: gov Identifier: NCT04797988.

关键词

MedicineAdverse effectClearanceUltrasoundVisual analogue scalePatient satisfactionSurgeryNuclear medicineUrologyRadiology

相关论文

查看 OTHER 分类全部论文