首页 /研究 /Design and Test of an In-Vivo Robotic Camera Integrated with Optimized Illumination System for Single-port Laparoscopic Surgery
SURGICAL

Design and Test of an In-Vivo Robotic Camera Integrated with Optimized Illumination System for Single-port Laparoscopic Surgery

Xiaolong Liu, Reza Yazdanpanah, Tao Zuo, Yong Guan, Gregory J. Mancini, Jindong Tan

发表年份
2018
引用次数
3

摘要

This paper proposes a novel in-vivo robotic laparo-scopic camera design with an optimized illumination system, which is a crucial component for achieving high imaging quality. The robotic camera design with three extendable wings can reserve sufficient on-board space to harbor the optimized illumination system without affecting the compactness of the camera. We contribute a freeform optical lens design method and develop three miniature optical lenses for the LEDs to achieve greater than 95% illumination uniformity, greater than 14, 000 lx illuminance on a target plane with a distance of 100 mm, and greater than 89% optical efficiency. The prototype is implemented and experimentally tested, which demonstrates great performance of the in-vivo robotic laparoscopic camera and the significance of the optimized illumination system.

关键词

IlluminanceLens (geology)Computer visionComputer scienceArtificial intelligenceCamera moduleOpticsPhysics