Effect of Magnetic Resonance Imaging on the Motion Accuracy of Magnetic Resonance Imaging-compatible Focused Ultrasound Robotic System

Anastasia Antoniou, Nikolas Evripidou, Antreas Chrysanthou, Leonidas Georgiou, Cleanthis Ioannides, Kyriakos Spanoudes, Christakis Damianou

发表年份
2024
引用次数
4

摘要

Purpose: The current study provides insights into the challenges of safely operating a magnetic resonance imaging (MRI)-guided focused ultrasound (MRgFUS) robotic system in a high-field MRI scanner in terms of robotic motion accuracy. Materials and Methods: Grid sonications were carried out in phantoms and excised porcine tissue in a 3T MRI scanner using an existing MRgFUS robotic system. Fast low-angle shot-based magnetic resonance thermometry was employed for the intraprocedural monitoring of thermal distribution. Results: Strong shifting of the heated spots from the intended points was observed owing to electromagnetic interference (EMI)-induced malfunctions in system's operation. Increasing the slice thickness of the thermometry sequence to at least 8 mm was proven an efficient method for preserving the robotic motion accuracy. Conclusions: These findings raise awareness about EMI effects on the motion accuracy of MRgFUS robotic devices and how they can be mitigated by employing suitable thermometry parameters.

关键词

Magnetic resonance imagingUltrasoundNuclear magnetic resonancePhysicsMedicineRadiology