Investigation of active tracking for robotic arm assisted magnetic resonance guided focused ultrasound ablation
Xu Xiao, Zhihong Huang, Martin A. Rube, Andreas Melzer
- 发表年份
- 2016
- 引用次数
- 9
- 访问权限
- 开放获取
摘要
BACKGROUND: Focused ultrasound surgery (FUS) is a technique that does not need invasive access to the patient while allowing precise targeted therapy. Magnetic resonance (MR) guided FUS provides capabilities for monitoring treatments. Because the targeted tumours are distributed at different positions, focus repositioning becomes necessary. METHODS: An MR compatible robot was used to increase the operational range of FUS application. Active tracking was developed to detect the robotic arm with regard to the MR coordinate system. The purpose of this study was to construct active tracking to allow a wide spatial range of repositioning the FUS transducer that is fast and accurate. The technique was characterised and validated by a series of positioning tests to prove its efficiency for guiding the robot. RESULTS: In the calibration range, tracking achieved an RMS accuracy of 0.63 mm. Results of phantom ablation showed a focal scanning precision Δx = 0.4 ± 0.37 mm, Δy = 0.4 ± 0.28 mm and Δz = 0.7 ± 0.66 mm. CONCLUSIONS: The active tracking localisation can be considered a feasible approach for the MR guided FUS system positioned by a robot. Copyright © 2016 John Wiley & Sons, Ltd.