Validation of dosimetric field matching accuracy from proton therapy using a robotic patient positioning system

Jonathan B. Farr, Avril O’Ryan-Blair, F. Jesseph, W Hsi, C. Allgower, Anthony Mascia, Allan F. Thornton, A. N. Schreuder

发表年份
2010
引用次数
9
访问权限
开放获取

摘要

Large area, shallow fields are well suited to proton therapy. However, due to beam production limitations, such volumes typically require multiple matched fields. This is problematic due to the relatively narrow beam penumbra at shallow depths compared to electron and photon beams. Therefore, highly accurate dose planning and delivery is required. As the dose delivery includes shifting the patient for matched fields, accuracy at the 1-2 millimeter level in patient positioning is also required. This study investigates the dosimetric accuracy of such proton field matching by an innovative robotic patient positioner system (RPPS). The dosimetric comparisons were made between treatment planning system calculations, radiographic film and ionization chamber measurements. The results indicated good agreement amongst the methods and suggest that proton field matching by a RPPS is accurate and efficient.