Inappropriate elevation of bispectral index values in robot assisted thyroidectomy with electromyographic endotracheal tube -A case report-
Mae-Hwa Kang, Chi-Bum In, Manho Kim, Kyoung-Ji Lim, Eun Young Park, Hyo Min Lee, Soo Kyung Lee
- 发表年份
- 2011
- 引用次数
- 9
- 访问权限
- 开放获取
摘要
We report a case of an erroneously elevated bispectral index (BIS) during robot assisted thyroidectomy using an electromyographic endotracheal tube (EMG tube), which is safe and useful for laryngeal electromyographic monitoring. Ten minutes after start of the operation, a sudden increase of BIS value up to 98 was noticed. The BIS values were not decreased to < 65 with supplemental anesthetics. The anesthetic method was changed from total intravenous anesthesia to balanced anesthesia. The BIS sensor and monitor were changed and other models were used. These interventions did not alter BIS values. BIS levels remained between 60 and 70 throughout the main procedure and intermittently increased to the mid-90s without any trace of poor signal quality. At the end of the surgery, the BIS values returned to normal range. The patient did not complain of intraoperative recall. Knowledge of potential interference from the use of an EMG tube must be considered when interpreting BIS.
关键词
相关论文
Robots and Jobs: Evidence from US Labor Markets
Daron Acemoğlu, Pascual Restrepo
2019
Reach and grasp by people with tetraplegia using a neurally controlled robotic arm
Leigh R. Hochberg, Daniel Bacher, Beata Jarosiewicz 等 11 位作者
2012
Campbell-Walsh urology
Alan J. Wein editor-in-chief
2012
Stroke rehabilitation
Peter Langhorne, Julie Bernhardt, Gert Kwakkel
2011