Multichannel EMG data acquisition system: Design and temporal analysis during human ankle joint movements
Maged S. Al-Quraishi, Asnor J. Ishak, Siti Anom Ahmad, M. K. Hasan
- 发表年份
- 2014
- 引用次数
- 12
摘要
This paper presents the implementation of four channel Electromyography (EMG) signal acquisition system for acquiring the EMG signal of the lower limb muscles during ankle joint movements. Furthermore, some post processing and statistical analysis for the recorded signal were presented. Four channels were implemented using instrumentation amplifier (INA114) for pre-amplification stage then the amplified signal subjected to the band pass filter to eliminate the unwanted signals. Operational amplifier (OPA2604) was involved for the main amplification stage to get the output signal in volts. The EMG signals were detected during movement of the ankle joint of a healthy subject. Then the signal was sampled at the rate of 2 kHz using NI6009 DAQ and Labview used for displaying and storing the acquired signal. For EMG temporal representation, mean absolute value (MAV) analysis algorithm is used to investigate the level of the muscles activity. This data will be used in future as a control input signal to drive the ankle joint exoskeleton robot.