Small deviations between planned and performed bone cuts using a CT‐based robotic‐arm‐assisted total knee arthroplasty system
Francesco Zambianchi, Nikita Matveitchouk, Marco Pavesi, Sebastiano Clemenza, Riccardo Cuoghi Costantini, Andrea Marcovigi, Stefano Seracchioli, Fabio Catani
- 发表年份
- 2024
- 引用次数
- 13
摘要
PURPOSE: Computed tomography (CT)-based robotic system for total knee arthroplasty (TKA) has shown improved accuracy compared to conventional. This study was designed to (1) confirm the accuracy of the robotic system in achieving the plan and (2) establish the alignment and positioning deviation between final components and planning, by measuring the discrepancy between final implant alignment and the corresponding planned cut. METHODS: Ninety-six cementless robotic-arm assisted (RA) TKAs were assessed. Bone resections were performed using the haptically controlled robotic arm. Alignment in the coronal and sagittal plane and resection depth of the distal femoral and proximal tibial cuts were recorded with a navigation planar probe. After final components were impacted, the probe was positioned on each implant surface to determine its alignment and positioning. RESULTS: The mean tibial resections and implanted tibial component's positioning were 0.4 mm (standard deviation, SD: 0.6) and 0.9 mm (SD: 0.8), respectively, higher than planned (p < 0.01). The tibial sagittal cut had 19/96 cases (19.8%) of ±1° outliers from plan. In 40/96 cases (41.7%), the tibial component was more prominent than planned of more than 1 mm. The mean femoral resections and impacted femoral component's positioning was 0.1 mm (SD: 0.8) and 0.2 mm (SD: 0.7), respectively, higher than planned. In 23/96 cases (24.0%), the femoral sagittal cut and femoral component coronal alignment deviated more than ±1° from plan. CONCLUSIONS: The computed tomography-based robotic-assisted TKA system showed good accuracy regarding bone preparation and component's positioning relative to the planning. Cementless tibial component impaction resulted in the most deviation from plan, with a large proportion of cases resulting in being more prominent than planned. LEVEL OF EVIDENCE: Level III.
关键词
相关论文
Robots and Jobs: Evidence from US Labor Markets
Daron Acemoğlu, Pascual Restrepo
2019
The Uncanny Valley [From the Field]
Masahiro Mori, Karl F. MacDorman, Norri Kageki
2012
Measurement Instruments for the Anthropomorphism, Animacy, Likeability, Perceived Intelligence, and Perceived Safety of Robots
Christoph Bartneck, Dana Kulić, Elizabeth A. Croft 等 4 位作者
2008
Reach and grasp by people with tetraplegia using a neurally controlled robotic arm
Leigh R. Hochberg, Daniel Bacher, Beata Jarosiewicz 等 11 位作者
2012