A preliminary study on using a Robotically-Actuated Delivery Sheath (RADS) for transapical aortic valve implantation

Gustaaf J. Vrooijink, Tim T. M. Ellenbroek, Paul Breedveld, Jan G. Grandjean, Sarthak Misra

发表年份
2014
引用次数
17

摘要

Recent technological advancements in cardiovascular surgery such as transapical transcatheter aortic valve implantation (TA-TAVI) enabled treatment to elderly that were initially declined surgery. However, valve malpositioning during TA-TAVI have been reported in several cases. In this preliminary study, we present a novel approach in which a Robotically-Actuated Delivery Sheath (RADS) is used to potentially facilitate valve positioning. A model is developed that describes the shape and articulating tip position of the RADS. We developed a two-dimensional ultrasound tracking method that evaluates the tip position of the RADS in ultrasound images. Both modeling and ultrasound tracking are combined into an integrated system that facilitates closed-loop control of the articulating tip of the RADS. Experiments are performed in order to evaluate the tracking accuracy of the RADS. Experiments show mean positioning errors of approximately 2 mm along the x- and y-axes. Our study demonstrates that the RADS can potentially provide compensation for beating heart and respiratory motions during valve positioning and deployment in TA-TAVI.