A medical robot for needle placement therapy in liver cancer
Xingguang Duan, Gui‐Bin Bian, Honghua Zhao, Xingtao Wang, Qiang Huang
- 发表年份
- 2010
- 引用次数
- 18
摘要
Hepatocellular carcinoma (HCC) is the second most common malignancy in China. As microwave ablation (MWA) is an effective method for liver cancer, a robotic surgical system with ultrasound-directed was designed to assist surgeons on positioning the needles. This robotic system includes a surgical robot with 5 degrees of freedom, a workstation for path-planning and image processing, a conventional 2D ultrasound device, and an electromagnetic (EM) tracking system. Surgery space, clinical operation requirements and optimal mechanical structure are the key factors to be considered in designing a medical robot suitable for use by surgeons. Based on the mechanics of the needle placement robot, we have conducted detailed kinematic analysis, including a combined numerical algorithm and coordinate mapping. Finally, the feasibility of the needle placement robot has been validated by experiment.