Robotic Ultrasound for Catheter Navigation in Endovascular Procedures
Fernanda Langsch, Salvatore Virga, Javier Esteban, Rüdiger Göbl, Nassir Navab
- 发表年份
- 2019
- 引用次数
- 28
摘要
Endovascular procedures require real time visual feedback on the location of inserted catheters. This is currently achieved using X-ray fluoroscopy, which causes exposure to radiation. This study describes an alternative method using a robotic ultrasound system for catheter tracking and navigation in endovascular interventions, focusing on endovascular aneurysm repair. This approach relies on the registration of pre-operative images to provide both a tracking trajectory and visual feedback of the real-time catheter position. The procedure was validated on healthy volunteers and on a phantom that included a realistic vessel structure, showing an average tracking error of the moving catheter tip of 1.78±1.02 mm.