Fetal tele‐echography using a robotic arm and a satellite link
Philippe Arbeille, José Luis Ausín Ruiz, P Hervé, Muriel Chevillot, Gérard Poisson, F. Perrotin
- 发表年份
- 2005
- 引用次数
- 69
- 访问权限
- 开放获取
摘要
OBJECTIVE: To design a method for conducting fetal ultrasound examinations in isolated hospital sites using a dedicated remotely controlled robotic arm (tele-echography). METHODS: Tele-echography was performed from our hospital (expert center) on 29 pregnant women in an isolated maternity hospital (patient site) 1700 km away, and findings were compared with those of conventional ultrasound examinations. At the patient site, a robotic arm holding the real ultrasound probe was placed on the patient's abdomen by an assistant with no experience of performing ultrasound. The robotic arm, remotely controlled with a fictive (expert) probe, reproduced the exact movements (tilting and rotating) of the expert hand on the real ultrasound probe. RESULTS: In 93.1% of the cases, all biometric parameters, placental location and amniotic fluid volume, were correctly assessed using the teleoperated robotic arm. In two cases, femur length could not be correctly measured. The mean duration of fetal ultrasound examination was 14 min (range, 10-18) and 18 min (range, 13-23) by conventional and tele-echography methods, respectively. The mean number of times the robotic arm was repositioned on the patient's abdomen was seven (range, 5-9). CONCLUSION: Tele-echography using a robotic arm provides the main information needed to assess fetal growth and the intrauterine environment within a limited period of time.
关键词
相关论文
Statistical Learning Theory
Yuhai Wu, Vladimir Vapnik
1999
Artificial intelligence: a modern approach
1995
Fractional Differential Equations
Igor Podlubný
2025
Applied Nonlinear Control
Jean-Jacques Slotine, Weiping Li
1991