Inadvertent Intraoperative Defibrillation Secondary to Electrocautery Grounding Pad Placement
Joseph A. McGuire, J.W. Awori Hayanga, Charles Barry, Anna E. Carpenter, Benjamin M. Frye, James P. Hughes, David Schwartzman, Heather K. Hayanga
- Year
- 2022
- Citations
- 2
- Access
- Open access
Abstract
Intraoperative defibrillation secondary to the usage of electrocautery in a patient with a cardiovascular implantable electronic device is a rare occurrence, and below-the-umbilicus electrocautery use causing inadvertent defibrillation is a near-zero risk. Defibrillation secondary to electrodispersive pad (EDP) radiofrequency dispersion has only ever been theorized. In this report, we describe the case of a 67-year-old male with an automatic implantable cardioverter defibrillator (AICD) undergoing a robotic-assisted left anterior total hip arthroplasty for left hip osteoarthritis who experienced inadvertent intraoperative defibrillation concurrent with electrocautery usage. The defibrillations ceased following contralateral EDP repositioning and application of a donut magnet overlying the patient's AICD.
Keywords
Related papers
Statistical Learning Theory
Yuhai Wu, Vladimir Vapnik
1999
Fractional Differential Equations
Igor Podlubný
2025
Applied Nonlinear Control
Jean-Jacques Slotine, Weiping Li
1991
Genetic Programming: On the Programming of Computers by Means of Natural Selection
John R. Koza
1992