Home /Research /Inadvertent Intraoperative Defibrillation Secondary to Electrocautery Grounding Pad Placement
OTHER

Inadvertent Intraoperative Defibrillation Secondary to Electrocautery Grounding Pad Placement

Joseph A. McGuire, J.W. Awori Hayanga, Charles Barry, Anna E. Carpenter, Benjamin M. Frye, James P. Hughes, David Schwartzman, Heather K. Hayanga

Year
2022
Citations
2
Access
Open access

Abstract

Intraoperative defibrillation secondary to the usage of electrocautery in a patient with a cardiovascular implantable electronic device is a rare occurrence, and below-the-umbilicus electrocautery use causing inadvertent defibrillation is a near-zero risk. Defibrillation secondary to electrodispersive pad (EDP) radiofrequency dispersion has only ever been theorized. In this report, we describe the case of a 67-year-old male with an automatic implantable cardioverter defibrillator (AICD) undergoing a robotic-assisted left anterior total hip arthroplasty for left hip osteoarthritis who experienced inadvertent intraoperative defibrillation concurrent with electrocautery usage. The defibrillations ceased following contralateral EDP repositioning and application of a donut magnet overlying the patient's AICD.

Keywords

MedicineDefibrillationImplantable cardioverter-defibrillatorSurgeryAnesthesiaCardiology