Home /Research /The development of a robotic approach to therapeutic ultrasound
OTHER

The development of a robotic approach to therapeutic ultrasound

Zhen Qiu, Chunhua Xie, S. Cochran, George Corner, Zhihong Huang, Changhui Song, S Daglish

Year
2009
Citations
5
Access
Open access

Abstract

Despite the recent wealth of research into therapy with high intensity focused ultrasound (HIFU), its clinical application still has limitations. These relate to the accuracy of the HIFU beam, the ability to visualize the target volume, and the understanding of the beam interaction with tissue. In the work reported in this paper, the output characteristics of a single channel HIFU transducer have been investigated in detail with the aid of a six-axis robot. Results show that by altering the drive parameters of the transducer clearly defined thermal or mechanical damage can be produced. The nature and patterns of damage produced by preprogrammed treatment volumes are now being investigated in tissue.

Keywords

TransducerHigh-intensity focused ultrasoundUltrasoundComputer scienceTherapeutic ultrasoundFocused ultrasoundBiomedical engineeringBeam (structure)AcousticsMedicine