Quantification of Force and Torque Applied by a High-Field Magnetic Resonance Imaging System on an Ultrasonic Motor for MRI-Guided Robot-Assisted Interventions
Peyman Shokrollahi, James M. Drake, A.A. Goldenberg
- Year
- 2017
- Citations
- 5
- Access
- Open access
Abstract
The risk of accidental dislodgement of robot-operated surgical mechanisms can lead to morbidity or mortality. The force and torque applied by a 3.0-tesla scanner on an ultrasonic motor are not fully known. The force and torque may displace the motor, which is not fully magnetic resonance imaging (MRI)-compatible but can be safely used in MR environments. A suspension apparatus was designed to measure the angles of deflection and rotation applied to the motor by MR magnetic fields. Three orientations and two power states of the motor were assessed inside the MR bore. The displacement force and torque were measured at eight locations with respect to the bore. The displacement force on the motor from 10 cm outside the magnet bore to 20 cm inside the bore ranged from 3 to 7 gF. The experimental measurements are in agreement with the theoretical values. Running the motor altered the force by 1 gF. The force does not significantly change when the MRI scanner is on. Considerable displacement force is applied to the motor, and no deflection torque is observed. Quantified values can be used to solve dynamic equations for robotic mechanisms intended for MRI-guided operations.
Keywords
Related papers
Applied Nonlinear Control
Jean-Jacques Slotine, Weiping Li
1991
Fractional Brownian Motions, Fractional Noises and Applications
Benoît B. Mandelbrot, John W. Van Ness
1968
Sliding Mode Control in Electro-Mechanical Systems
Vadim Utkin, Jürgen Guldner, Jingxin Shi
2010
Hybrid Position/Force Control of Manipulators
Marc H. Raibert, John Craig
1981