Development and Feasibility of a Robotic Laparoscopic Clipping Tool for Wound Closure and Anastomosis
Axel Krieger, Justin D. Opfermann, Peter C.W. Kim
- Year
- 2017
- Citations
- 5
- Access
- Open access
Abstract
This paper reports the design, development, and initial evaluation of a robotic laparoscopic clipping tool for single manipulator wound closure and anastomosis (tubular reconnection). The tool deploys biodegradable clips and clasps with the goal of (i) integrating grasping and suturing into a single device for single hand or manipulator use, (ii) applying the equivalent of interrupted sutures without the need of managing suture thread, and (iii) allowing for full six degrees-of-freedom (DOFs) laparoscopic control when mounted on a robot arm. The specifications, workflow, and detailed design of the robotic laparoscopic tool and injection molded bio-absorbable T shaped clip and locking clasp are reported. The clipping tool integrates forceps to grab and stabilize tissue and a clip and clasp applier to approximate and fixate the tissue. A curved needle is advanced on a circular needle path and picks up and drags clips through tissue. The clip is then tightened through the tissue and a clasp is clamped around the clip, before the clip is released from the needle. Results of several bench test runs of the tool show: (a) repeatable circular needle drive, (b) successful pick-up and deployment of clips, (c) successful shear of the clip to release the clip from the needle, and (d) closure of clasp on clip with an average of 2.0 N holding force. These data indicate that the robotic laparoscopic clipping tool could be used for laparoscopic wound closure and anastomosis.
Keywords
Related papers
Statistical Learning Theory
Yuhai Wu, Vladimir Vapnik
1999
Artificial intelligence: a modern approach
1995
Applied Nonlinear Control
Jean-Jacques Slotine, Weiping Li
1991
A new optimizer using particle swarm theory
R.C. Eberhart, James Kennedy
2002