Preliminary results on OCT-based position control of a concentric tube robot

Y. Baran, Kanty Rabenorosoa, Guillaume J. Laurent, P. Rougeot, Nicolas Andreff, Brahim Tamadazte

Year
2017
Citations
17

Abstract

This paper presents preliminary results on OCT (Optical Coherence Tomography) based control of a concentric tube robot (CTR). The potential use of OCT system is presented then its ability to perform an optical biopsy and an eye-to-hand configuration for robot control is highlighted. A prototype of CTR with three curved tubes is shortly described then a model-less OCT-based controller is detailed. The extraction of visual features from OCT images is performed by normalized cross-correlation. The proposed control scheme was experimentally validated. The positioning accuracy of CTR control was evaluated in benchtop trials. The error in tip placement can reach 10 μm along X and 74 μm along Y according to controller parameters. The 2D positioning (XY plan) is also tested then a final error about 55 μm is achieved.