Robot-Assisted Aortic Valve Replacement Using a Novel Sutureless Bovine Pericardial Prosthesis Proof of Concept as an Alternative to Percutaneous Implantation
Rakesh M. Suri, Harold M. Burkhart, Hartzell V. Schaff
- Year
- 2010
- Citations
- 21
Abstract
Objective Percutaneous aortic valve implantation within native valve calcium has progressed to clinical use despite the absence of data proving equivalence to complete surgical excision and prosthetic valve replacement. A novel self-expanding sutureless bovine pericardial prosthesis (Sorin Perceval) derived from a proven stented valve has been successfully used in humans recently through an open transaortic approach. We sought to develop a minimally invasive technique for native aortic valve excision and sutureless prosthetic aortic valve replacement using robot assistance. Methods The da Vinci S-HD system was used to open and suspend the pericardium anterior to the phrenic nerve in cadavers. A transthoracic cross-clamp was placed across the midascending aorta, following which a transverse aortotomy was made. The native aortic valve cusps were excised, and annular calcium was removed with robotic instruments. After placement of three guide sutures, the Perceval self-expanding pericardial prosthesis mounted on a flexible delivery system was inserted through a working port and lowered into the aortic annulus. Results Successful implantation of all valves was possible using a 3-cm right second intercostal space working port, along with two additional 1-cm instrument ports. A standard transverse aortotomy was sufficient for examination/debridement of the native aortic valve cusps, sizing of the annulus, and deployment of the nitinol-stented, bovine pericardial prosthesis. Delivery, seating, and stability of the device were easily confirmed above and below the aortic valve annulus using the robotic camera. Conclusions Complete excision of diseased native aortic valve cusps with robot assistance facilitates accurate and reproducible aortic valve replacement using a novel self-expanding sutureless version of a proven bovine pericardial prosthesis. This approach is comparable to the current surgical gold standard and is ready for clinical use as an alternative to percutaneous aortic valve implantation.
Keywords
Related papers
Robots and Jobs: Evidence from US Labor Markets
Daron Acemoğlu, Pascual Restrepo
2019
Reach and grasp by people with tetraplegia using a neurally controlled robotic arm
Leigh R. Hochberg, Daniel Bacher, Beata Jarosiewicz +8 more
2012
Campbell-Walsh urology
Alan J. Wein editor-in-chief
2012
Stroke rehabilitation
Peter Langhorne, Julie Bernhardt, Gert Kwakkel
2011