Home /Research /Commissioning and initial experience with the first clinical gantry‐mounted proton therapy system
OTHER

Commissioning and initial experience with the first clinical gantry‐mounted proton therapy system

Tianyu Zhao, Baozhou Sun, K Grantham, Leith Rankine, Bin Cai, Sreekrishna Goddu, Lakshmi Santanam, Nels Knutson, Tiezhi Zhang, Michael Reilly, Beth Bottani, Jeffrey D. Bradley, Sasa Mutic, Eric Klein

Year
2016
Citations
36
Access
Open access

Abstract

The purpose of this study is to describe the comprehensive commissioning process and initial clinical experience of the Mevion S250 proton therapy system, a gantry-mounted, single-room proton therapy platform clinically implemented in the S. Lee Kling Proton Therapy Center at Barnes-Jewish Hospital in St. Louis, MO, USA. The Mevion S250 system integrates a compact synchrocyclotron with a C-inner gantry, an image guidance system and a 6D robotic couch into a beam delivery platform. We present our commissioning process and initial clinical experience, including i) CT calibration; ii) beam data acquisition and machine characteristics; iii) dosimetric commissioning of the treatment planning system; iv) validation through the Imaging and Radiation Oncology Core credentialing process, including irradiations on the spine, prostate, brain, and lung phantoms; v) evaluation of localization accuracy of the image guidance system; and vi) initial clinical experience. Clinically, the system operates well and has provided an excellent platform for the treatment of diseases with protons.

Keywords

Proton therapyMedical physicsRadiation treatment planningCredentialingNuclear medicineMedicineProject commissioningDelivery systemRadiation therapyRadiology

Related papers

Browse all OTHER papers