Safety in Senhance™ Robotic Gastrointestinal Surgery in 530 Patients
Ludger Staib, Clemens Poth, Florian Schilcher, Olaf Hansen, Vivianda Menke, Dietmar Stephan, Frank Willeke
- 发表年份
- 2023
- 引用次数
- 3
摘要
The Senhance Robotic System™ (Asensus Surgical, Durham, NC, USA) has been used in abdominal surgery since 2016, and provides an eye-tracker for camera movement and haptic tactile feedback. Safety aspects are very important in robotic surgery, such as regarding the presence of system malfunctions and surgical outcomes. The data for robotic function in gastrointestinal surgical procedures in 530 patients (colorectal surgery, fundoplication, others) were prospectively listed in the TRUST registry after informed patient consent in three German gastrointestinal surgery centers (center A, N = 46 patients; center B, N = 457; center C, N =27). Adverse events were noted in 14.3% (76/530 patients) of the overall surgeries, with an equal distribution among the procedures. Robotic malfunctions, such as console/camera/arm malfunctions, collisions, or limited motion, were experienced in 5.5 % (29/530 patients), with some differences among the centers (A, 0.0%; B, 4.2%; C, 37%). These differences were explained in terms of team experience and case load. In conclusion, the Senhance™ Robotic System can be safely applied to routine abdominal surgery procedures.