Flexible positioning of a large area detector using an industrial robot
Christina Reinhard, Michael Drakopoulos, Christopher M. Charlesworth, Andrew James, Hiten Patel, Paul Tutthill, Davide Crivelli, Hans Deyhle, Sharif Ahmed
- 发表年份
- 2022
- 引用次数
- 4
- 访问权限
- 开放获取
摘要
The DIAD beamline for Dual Imaging and Diffraction at Diamond Light Source has opted to use an industrial robot to position its Dectris Pilatus 2M CdTe diffraction detector. This setup was chosen to enable flexible positioning of the detector in a quarter-sphere around the sample position whilst reliably holding the large weight of 139 kg of detector, detector mount and cabling in a stable position. Metrology measurements showed that the detector can be positioned with a linear repeatability of <19.7 µm and a rotational repeatability of <16.3 µrad. The detector position stays stable for a 12 h period with <10.1 µm of movement for linear displacement and <3.8 µrad for rotational displacement. X-ray diffraction from calibration samples confirmed that the robot is sufficiently stable to resolve lattice d-spacings within the instrumental broadening given by detector position and beam divergence.
关键词
相关论文
Statistical Learning Theory
Yuhai Wu, Vladimir Vapnik
1999
Fractional Differential Equations
Igor Podlubný
2025
Applied Nonlinear Control
Jean-Jacques Slotine, Weiping Li
1991
Genetic Programming: On the Programming of Computers by Means of Natural Selection
John R. Koza
1992