首页 /研究 /[Robot-guided percutaneous kyphoplasty in treatment of multi-segmental osteoporotic vertebral compression fracture].
OTHER

[Robot-guided percutaneous kyphoplasty in treatment of multi-segmental osteoporotic vertebral compression fracture].

Shu Lin, Jiang Hu, Lun Wan, Liuyi Tang, Yue Wang, Yang Yu, Wei Zhang

发表年份
2020
引用次数
7

摘要

OBJECTIVE: To evaluate the safety and effectiveness of robot-guided percutaneous kyphoplasty (PKP) in treatment of multi-segmental thoracolumbar osteoporotic vertebral compression fracture (OVCF). METHODS: >0.05). The time to establish the tunnel, the times of fluoroscopy, the dose of fluoroscopy, the deviation of puncture, the distribution of bone cement, the leakage of bone cement, the puncture angle, and the postoperative VAS score, midline vertebral height, and Cobb angle were recorded and compared. RESULTS: >0.05). CONCLUSION: Robot-guided PKP in treatment of multi-segmental thoracolumbar OVCF can shorten the operation time, improve the accuracy of puncture, reduce the times and dose of fluoroscopy, reduce the leakage of bone cement, and achieve better cement distribution.

关键词

FluoroscopyMedicineCobb angleVertebral compression fracturePercutaneousVisual analogue scaleBone cementOsteoporosisSurgeryRadiology