An Ultrasound Imaging-Guided Robotic HIFU Ablation Experimental System and Accuracy Evaluations

Chih Yu An, Jia Hao Syu, Ching‐Shiow Tseng, Chih‐Ju Chang

发表年份
2017
引用次数
15
访问权限
开放获取

摘要

In recent years, noninvasive thermal treatment by using high-intensity focused ultrasound (HIFU) has high potential in tumor treatment. The goal of this research is to develop an ultrasound imaging-guided robotic HIFU ablation system for tumor treatment. The system integrates the technologies of ultrasound image-assisted guidance, robotic positioning control, and HIFU treatment planning. With the assistance of ultrasound image guidance technology, the tumor size and location can be determined from ultrasound images as well as the robotic arm can be controlled to position the HIFU transducer to focus on the target tumor. After the development of the system, several experiments were conducted to measure the positioning accuracy of this system. The results show that the average positioning error is 1.01 mm with a standard deviation 0.34, and HIFU ablation accuracy is 1.32 mm with a standard deviation 0.58, which means this system is confirmed with its possibility and accuracy.

关键词

UltrasoundHigh-intensity focused ultrasoundAblationThermal ablationPositioning systemComputer scienceBiomedical engineeringFocused ultrasoundFocus (optics)Medical physics