Safe and effective use of active robotics for TKA: Early results of a multicenter study
Bernard N. Stulberg, Jayson D. Zadzilka, Stefan Kreuzer, Yair Kissin, Ralph Liebelt, William J. Long, Valentina Campanelli
- 发表年份
- 2021
- 引用次数
- 19
摘要
BACKGROUND: A novel active robotic system for total knee arthroplasty (TKA) performs automated milling of bone surfaces. Study objectives were to assess system safety and effectiveness in a US population. METHODS: A multicenter clinical trial was conducted, following 115 patients for at least 6-months. A pre-defined list of robot-related adverse events was used to evaluate safety. Efficacy was assessed radiographically comparing planned versus achieved coronal limb alignment. RESULTS: No pre-defined adverse events occurred and postoperative limb alignment more than ±3° from plan occurred in 11.2 % of cases. CONCLUSION: Active robotics for TKA is safe and effective as demonstrated in this trial.