A Robotic Ankle–Foot Prosthesis With Active Alignment
Andrew K. LaPrè, Brian R. Umberger, Frank C. Sup
- 发表年份
- 2016
- 引用次数
- 23
摘要
An ankle–foot prosthesis designed to mimic the missing physiological limb generates a large sagittal moment during push off which must be transferred to the residual limb through the socket connection. The large moment is correlated with high internal socket pressures that are often a source of discomfort for the person with amputation, limiting prosthesis use. In this paper, the concept of active alignment is developed. Active alignment realigns the affected residual limb toward the center of pressure (CoP) during stance. During gait, the prosthesis configuration changes to shorten the moment arm between the ground reaction force (GRF) and the residual limb. This reduces the peak moment transferred through the socket interface during late stance. A tethered robotic ankle prosthesis has been developed, and evaluation results are presented for active alignment during normal walking in a laboratory setting. Preliminary testing was performed with a subject without amputation walking with able-bodied adapters at a constant speed. The results show a 33% reduction in the peak resultant moment transferred at the socket limb interface.