A Simple Method for Validation and Verification of Pipettes Mounted on Automated Liquid Handlers

Michael Stangegaard, Anders J. Hansen, Tobias Guldberg Frøslev, Niels Morling

发表年份
2010
引用次数
29
访问权限
开放获取

摘要

We have implemented a simple, inexpensive, and fast procedure for validation and verification of the performance of pipettes mounted on automated liquid handlers (ALHs) as necessary for laboratories accredited under ISO 17025. A six- or seven-step serial dilution of OrangeG was prepared in quadruplicates in a flat-bottom 96-well microtiter plate, manually using calibrated pipettes. Each pipette of the liquid handler (1-8) dispensed a selected volume (1-200 μL) of OrangeG eight times into the wells of the microtiter plate. All wells contained a total of 200 μL liquid. The absorbance was read, and the dispensed volume of each pipette was calculated based on a plot of volume and absorbance of a known set of OrangeG dilutions. Finally, the percent inaccuracy (%d) and the imprecision (%CV) of each pipette were calculated. Using predefined acceptance criteria, each pipette was then either approved or failed. Failed pipettes were either repaired or the volume deviation was compensated for by applying a calibration curve in the liquid-handler software. We have implemented the procedure on a Sias Xantus, an MWGt TheONYX, four Tecan Freedom EVO, a Biomek NX Span-8, and four Biomek 3000 robots, and the methods are freely available. In conclusion, we have set up a simple, inexpensive, and fast solution for the continuous validation of ALHs used for accredited work according to the ISO 17025 standard. The method is easy to use for aqueous solutions but requires a spectrophotometer that can read microtiter plates.

关键词

PipetteSimple (philosophy)Computer scienceBiomedical engineeringChromatographyEngineeringChemistry

相关论文

查看 OTHER 分类全部论文