首页 /研究 /Development of a robotic FD‐CT‐guided navigation system for needle placement—preliminary accuracy tests
MANIPULATION

Development of a robotic FD‐CT‐guided navigation system for needle placement—preliminary accuracy tests

Saúl Tovar‐Arriaga, Ralf Tita, Jesús Carlos Pedraza‐Ortega, Efrén Gorrostieta, Willi A. Kalender

发表年份
2011
引用次数
52

摘要

BACKGROUND: A needle placement system using a serial robot arm for manipulation of biopsy and/or treatment needles is introduced. A method for fast calibration of the robot and the preliminary accuracy tests of the robotic system are presented. METHODS: The setup consists of a DLR/KUKA Light Weight Robot III especially designed for safe human/robot interaction mounted on a mobile platform, a robot-driven angiographic C-arm system and a navigation system. RESULTS: Calibration of the robot with the navigation system has a residual error of 0.23 mm (rms) with a standard deviation of ± 0.1 mm. Needle targeting accuracy with different trajectories was 1.2 mm (rms) with a standard deviation of ± 0.4 mm. CONCLUSIONS: Robot absolute positioning accuracy was reduced to the navigation camera accuracy. The approach includes control strategies that may be very useful for interventional applications.

关键词

Navigation systemRobotComputer scienceCalibrationStandard deviationComputer visionRobotic armArtificial intelligenceAccuracy and precisionSimulation

相关论文

查看 MANIPULATION 分类全部论文