首页 /研究 /Achieving Accurate Ligament Balancing Using Robotic-Assisted Unicompartmental Knee Arthroplasty
SURGICAL

Achieving Accurate Ligament Balancing Using Robotic-Assisted Unicompartmental Knee Arthroplasty

Johannes F. Plate, Ali Mofidi, Sandeep Mannava, Beth P. Smith, Jason E. Lang, Gary G. Poehling, Michael A. Conditt, Riyaz H. Jinnah

发表年份
2013
引用次数
105
访问权限
开放获取

摘要

Unicompartmental knee arthroplasty (UKA) allows replacement of a single compartment in patients with limited disease. However, UKA is technically challenging and relies on accurate component positioning and restoration of natural knee kinematics. This study examined the accuracy of dynamic, real-time ligament balancing using a robotic-assisted UKA system. Surgical data obtained from the computer system were prospectively collected from 51 patients (52 knees) undergoing robotic-assisted medial UKA by a single surgeon. Dynamic ligament balancing of the knee was obtained under valgus stress prior to component implantation and then compared to final ligament balance with the components in place. Ligament balancing was accurate up to 0.53 mm compared to the preoperative plan, with 83% of cases within 1 mm at 0°, 30°, 60°, 90°, and 110° of flexion. Ligamentous laxity of 1.31 ± 0.13 mm at 30° of flexion was corrected successfully to 0.78 ± 0.17 mm (P < 0.05). Robotic-assisted UKA allows accurate and precise reproduction of a surgical balance plan using dynamic, real-time soft-tissue balancing to help restore natural knee kinematics, potentially improving implant survival and functional outcomes.

关键词

Unicompartmental knee arthroplastyMedicineValgusLigamentKinematicsSoft tissueArthroplastySurgeryOsteoarthritis

相关论文

查看 SURGICAL 分类全部论文